What do you do when you find an error in a certificate you already issued?

Quietly editing a document that is already in the customer's hands does not fix the error — it erases the trail. This is why ISO/IEC 17025 asks for a revision rather than a correction: the original stays in the record as it was, and the new version says what changed and why. For that to hold, the decision has to survive in the software too; if the edit button is still there, the rule lives only in the training slides. In SenSight QMS, opening an issued calibration record says so plainly: changing a result requires a new revision, and existing signatures become invalid.
Try it in the app
In SenSight QMS: opening an issued calibration record tells you that changing a result requires a new revision.
Watch this on our YouTube channel

SenSight QMS — quality management and calibration records in one place

SenSight QMS keeps the record side of ISO 9001 and ISO 17025 in one place - for accredited calibration and testing laboratories, and for manufacturers who run quality separately from their ERP.